• Critical care medicine · Oct 2005

    Clinical Trial

    Transcutaneous PCO2 monitoring in critically ill adults: clinical evaluation of a new sensor.

    • Karim Bendjelid, Nicolas Schütz, Martin Stotz, Isabelle Gerard, Peter M Suter, and Jacques-André Romand.
    • Surgical Intensive Care Unit, Department of Anesthesiology, Pharmacology and Surgical Intensive Care, Geneva University Hospitals, Geneva, Switzerland.
    • Crit. Care Med. 2005 Oct 1; 33 (10): 2203-6.

    ObjectiveIn critically ill patients, arterial blood gas analysis is the gold standard for evaluating systemic oxygenation and carbon dioxide partial pressure. A new miniaturized carbon dioxide tension Pco2-Spo2 single sensor (TOSCA, Linde Medical Sensors AG, Basel, Switzerland) continuously and noninvasively (transcutaneously) monitors both Paco2 and oxygen saturation by pulse oximetry (Spo2). The present study was designed to investigate the usability and the accuracy of this device in critically ill patients.DesignProspective clinical investigation.SettingA 20-bed, university-affiliated, surgical intensive care unit.PatientsPatients admitted after major surgery, multiple trauma, or septic shock equipped with an arterial catheter.InterventionsThe heated (42 degrees C) sensor was fixed at the earlobe using an attachment clip. Transcutaneous Pco2 (TcPco2) measurements were correlated with Paco2 values (measured using a blood gas analyzer). In addition, the differences between Paco2 and TcPco2 values were evaluated using the method of Bland-Altman.Measurements And Main ResultsWe studied 55 patients, aged 18-80 (mean 57 +/- 15) yrs. A total of 417 paired measurements were compared. Correlation between TcPco2 and Paco2 was r = .86 (p < .01) in the Paco2 range of 24-101 mm Hg. Mean bias (+/-sd) between the two methods of measurement (Bland-Altman analysis) was 1.2 +/- 6.0 mm Hg with TcPco2 slightly overestimating arterial carbon dioxide tension. Nineteen percent of the measured values were outside of the acceptable clinical range of agreement of +/-7.5 mm Hg.ConclusionsThe present study suggests that Paco2 can be acceptably assessed by measuring TcPco2 using the TOSCA Pco2-Spo2 sensor.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…