• J Shoulder Elbow Surg · May 2007

    Comparative Study

    A biomechanical comparison of locking plate and locking nail implants used for fractures of the proximal humerus.

    • Jeffrey Kitson, Grant Booth, and Robert Day.
    • Department of Orthopaedic Surgery, Royal Perth Hospital, Perth, Western Australia. kathandjeff@doctors.org.uk
    • J Shoulder Elbow Surg. 2007 May 1; 16 (3): 362-6.

    AbstractThe aim of this study was to determine the biomechanical behavior of 2 different implants used in the fixation of proximal humeral fractures. The 2 implants in this study are specifically designed for the fixation of proximal humeral fractures, and both are based on the concept of fixed-angle locking screws. A reproducible 3-part fracture was created in paired human cadaveric bone and then fixed via the locking screw implants. Stress/strain curves for the bone-implant construct were created for loads applied in different directions to determine the relative stiffness below the yield point. After this, each construct was tested to failure with a valgus bending load. The locking nail implant provided a significantly stiffer construct to valgus, extension, and flexion cantilever bending loads below the failure threshold. The valgus load to failure was significantly higher for the nail. The mode of failure was different between implants.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.