• J Extra Corpor Technol · Mar 2007

    Comparative Study

    Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery.

    • Kirti P Patel, Gary W Hay, Mahesh Keitheri Cheteri, and David W Holt.
    • Memorial Hermann Baptist Hospital, Beaumont, Texas, USA. patelpump@sbcglobal.net
    • J Extra Corpor Technol. 2007 Mar 1; 39 (1): 10-7.

    AbstractThe purpose of this study was to compare the variation in hemoglobin (Hgb) values among various point-of-care (POC) analyzers available on the market. Eight analyzers (Gem 3000, ABL 720, ABL 77, Rapidpoint 405, IL 682, GemOPL, Hb 201+, and manual/centrifugation) were compared with the Hgb values from the Beckman Coulter LH750. A total of 72 patient samples were analyzed on each test instrument. The samples were obtained after intubation, after heparinization, during cardiopulmonary bypass, and after protamine administration. Four of the samples were excluded from the study because of delayed sample analysis. The calculated mean differences of reference test method Hgb (mean +/- SD) for all samples (n = 68) were Gem 3000 = 1.431 +/- 0.396 g/dL; ABL 720 = -0.224 +/- 0.240 g/dL; ABL 77 = 0.341 +/- 0.578 g/dL; Rapidpoint 405 = 0.001 +/- 0.205 g/dL; IL 682 = -0.137 +/- 0.232 g/dL; GemOPL = 0.774 +/- 0.427 g/dL; Hb 201+ = 0.110 +/- 0.524 g/dL; and manual/ centrifugation = 0.547 +/- 0.499 g/dL. Cumulative results indicated that the bias in Hgb values from the Gem 3000, ABL720, ABL 77, IL 682, GemOPL, and the manual method were statistically significant (p < .05), compared with the Coulter LH750. Additionally, only the Rapidpoint 405 and Hb 201+ most closely matched the values from the Coulter LH750 (p > .05). Some of the methodologies have previously been shown to be affected during hemodilution, hypoproteinemia, and/or after blood transfusion. There is variability among methodologies, which can give rise to statistically different Hgb values, and one should consider the "ideal" instrument based on this and many other factors. Based on our results, the rank order of closest approximation to the Coulter LH750 measurement was Rapidpoint 405, Hb 201+, IL 682, ABL 720, ABL 77, manual/centrifugation, GemOPL, and Gem 3000.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.