• Thromb Haemostasis · Jul 2014

    Fibrinogen measurement in cardiac surgery with cardiopulmonary bypass: analysis of repeatability and agreement of Clauss method within and between six different laboratories.

    • C Solomon, E Baryshnikova, A Tripodi, C J Schlimp, H Schöchl, J Cadamuro, D Winstedt, L Asmis, and M Ranucci.
    • Associate Prof. Cristina Solomon, MD, Commercial Development Acquired Bleeding Disorders, CSL Behring GmBH, Emil-von-Behring-Strasse 76, 35041 Marburg, Germany, Tel: +49 6421 39 5813, Fax: +49 6421 39 4146, E-mail: Cristina.Solomon@cslbehring.com.
    • Thromb Haemostasis. 2014 Jul 3; 112 (1): 109-17.

    AbstractPlasma fibrinogen concentration is important for coagulopathy assessment, and is most commonly measured using the Clauss method. Several factors, including device type and reagent, have been shown to affect results. The study objective was to evaluate performance and repeatability of the Clauss method and to assess differences between measurements performed during and after cardiopulmonary bypass (CPB), by testing plasma samples from patients undergoing cardiac surgery with CPB. Samples were collected from 30 patients before surgery, approximately 20 minutes before weaning from CPB, and 5 minutes after CPB and protamine. Fibrinogen concentration was determined using the Clauss method at six quality-controlled specialised laboratories, according to accredited standard operating procedures. Regarding within-centre agreement for Clauss measurement, mean differences between duplicate measurements were between 0.00 g/l and 0.15 g/l, with intervals for 95% limits of agreement for mean Bland-Altman differences up to 1.3 g/l. Regarding between-centre agreement, some mean differences between pairs of centres were above 0.5 g/l. Differences of up to ~2 g/l were observed with individual samples. Increased variability was observed between centres, with inter-class correlation values below 0.5 suggesting only fair agreement. There were no significant differences in fibrinogen concentration before weaning from CPB and after CPB for most centres and methods. In conclusion, considerable differences exist between Clauss-based plasma fibrinogen measured using different detection methods. Nevertheless, the similarity between measurements shortly before weaning from CPB and after CPB within centres suggests that on-pump measurements could provide an early estimation of fibrinogen deficit after CPB and thus guidance for haemostatic therapy.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.