• J Clin Monit Comput · Jun 2017

    Randomized Controlled Trial

    Performance of a closed-loop feedback computer-controlled infusion system for maintaining blood pressure during spinal anaesthesia for caesarean section: a randomized controlled comparison of norepinephrine versus phenylephrine.

    • Ngan Kee Warwick D WD Department of Anaesthesia and Intensive Care, The Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, Hong Kong, China. warwick@cuhk., Kim S Khaw, Yuk-Ho Tam, Floria F Ng, and Shara W Lee.
    • Department of Anaesthesia and Intensive Care, The Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, Hong Kong, China. warwick@cuhk.edu.hk.
    • J Clin Monit Comput. 2017 Jun 1; 31 (3): 617-623.

    AbstractClosed-loop feedback computer-controlled vasopressor infusion has been previously described for maintaining blood pressure during spinal anaesthesia for caesarean section but there are limited data available comparing the relative performance of different vasopressors. The aim of this study was to compare the performance of norepinephrine versus phenylephrine in this system. Data from a randomized, two-arm parallel group, double-blinded controlled trial were reanalyzed. 104 patients scheduled for elective caesarean section under spinal anaesthesia were randomized to receive computer-controlled closed-loop infusion of either norepinephrine 5 µg ml-1 or phenylephrine 100 µg ml-1. This was started immediately after induction of spinal anaesthesia and used an algorithm designed to maintain systolic blood pressure near baseline until fetal delivery. Performance error calculations were used to compare the performance of the two vasopressors. The primary outcome was defined as the median absolute performance error. Median performance error, wobble and divergence were also compared. Median absolute performance error was smaller in the norepinephrine group (median 3.79 [interquartile range 2.82-5.17] %) versus the phenylephrine group (4.70 [3.23-6.57] %, P = 0.028). In addition, median performance error was smaller (0.75 [-1.56-2.52] %) versus 2.61 [0.83-4.57] %, P = 0.002) and wobble was smaller (2.85 [2.07-5.17] %) versus 3.39 [2.62-4.90] %, P = 0.028) in the norepinephrine group versus the phenylephrine group. Divergence was similar between groups. The precision of the control of blood pressure was greater with norepinephrine compared with phenylephrine at the drug concentrations used.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.