    A novel method of evaluation of three heat-moisture exchangers in six different ventilator settings.

    • N Unal, J K Kanhai, S L Buijk, J C Pompe, W P Holland, I Gültuna, C Ince, B Saygin, and H A Bruining.
    • Department of Anesthesiology and Reanimation, Medical Faculty, University of Ankara, Turkey.
    • Intensive Care Med. 1998 Feb 1; 24 (2): 138146138-46.

    ObjectiveThe purpose of this study was to assess and compare the humidification, heating, and resistance properties of three commercially available heat-moisture exchangers (HMEs). To mimic clinical conditions, a previously validated, new, realistic experimental set-up and measurement protocol was used.DesignProspective, comparative experimental study.SettingSurgical Intensive Care Unit, University Hospital of Rotterdam.MaterialsAn experimental set-up consisting of a patient model, measurement systems, and ventilator and three different HME types.InterventionsThe air flow, pressure in the ventilation circuit, pressure difference over the HME, and partial water vapour pressure and temperature at each side of the HMEs were measured. Measurements were repeated every 30 min during the first 2 h and every hour up to 24 h for each HME at six different ventilator settings. The mean inspiratory and maximum expiratory resistance, flow-weighted mean absolute humidity and temperature outputs, and humidification and heating efficiencies of HMEs were calculated.Measurements And ResultsThe Dar Hygroster had the highest humidity output, temperature output, humidification efficiency, and heating efficiency values throughout the study (32.8 +/- 21. mg/l, 32.2 +/- 0.8 degrees C, 86.3 +/- 2.3%, and 0.9 +/- 0.01%, respectively) in comparison to the Humid-Vent Filter (25.3 +/- 3.2 mg/l, 31.9 +/- 0.8 degrees C, 72.2 +/- 5.3%, 0.9 +/- 0.02%, respectively) and the Pall Ultipor BB100 breathing circuit filter (23.4 +/- 3 mg/l, 28.3 +/- 0.7 degrees C, 68.8 +/- 5.9%, 0.8 +/- 0.02%, respectively). The inspiratory and expiratory resistance of the HMEs remained below clinically acceptable maximum values (2.60 +/- 0.04 and 2.45 +/- 0.05 cmH2O/l per s, respectively).ConclusionThe Dar Hygroster filter was found to have the highest humidity and temperature output of all three HMEs, the Humid-Vent filter had a satisfactory humidity output only at low tidal volume flow rate and minute volume settings, whereas the Pall Ultipore BB 100 never achieved a sufficient humidity and temperature output.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…