    Bipolar vessel sealing: instrument contamination and wear have little effect on seal quality and success in a porcine in vitro model.

    • Christian W Wallwiener, Susanna H Junginger, Wolfgang Zubke, Sara Y Brucker, Markus D Enderle, Alexander Neugebauer, Birgitt Schönfisch, and Markus Wallwiener.
    • Department of Obstetrics and Gynecology, University of Tübingen, Calwerstr. 7, 72076, Tübingen, Germany, cwallwiener@googlemail.com.
    • Langenbecks Arch Surg. 2014 Oct 1; 399 (7): 863-71.

    PurposeThe aim of this study is to systematically investigate under standardized experimental conditions the effects of instrument contamination and wear on the quality of bipolar vessel sealing (BVS) achieved using a reusable instrument.MethodsThe study was designed as a prospective, randomized, and controlled in vitro study and conducted in an academic research environment. Reusable bipolar coagulation forceps (BiClamp® 200 C, ERBE Elektromedizin) were used to apply sealing pressures of 300-1,100 mN/mm(2) to 239 renal arteries from commercially slaughtered female pigs (Swabian-Hall Swine). Forceps jaws were coated with porcine blood, blood and collagen, or blood, collagen, and fat to simulate instrument contamination with biological material during surgery. Clinical wear was mimicked by sandpaper abrasion. The main outcome measures were seal success (resistance to 250 mmHg intraluminal pressure for 2 min) and seal stability (burst pressure).ResultsSealing pressure had a significant impact, with 800 mN/mm(2) producing the best sealing results. Seal success increased with total energy applied to the tissue, a higher maximum temperature, and longer coagulation as indicated by desiccation time. Experimental contamination had no significant impact on seal success and only a limited effect on seal stability. Similarly, abrasive wear also had no significant effect on either seal quality or seal strength.ConclusionsThe impact of bipolar forceps contamination and wear on seal success and quality was negligible in our in vitro model. To achieve high-quality seals, it is essential to use adequate sealing pressures. Our findings could have direct implications for the design and clinical handling of BVS instruments.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…