Journal of anesthesia · Aug 2016
Lateral deviation of four types of epidural catheters from the lumbar epidural space into the intervertebral foramen.
- Tetsuya Uchino, Masahiro Miura, Yoshimasa Oyama, Shigekiyo Matsumoto, Chihiro Shingu, and Takaaki Kitano.
- Department of Anesthesiology, Faculty of Medicine, Oita University, 1-1 Idaigaoka, Hazamacho, Yufushi, Oita, 879-5503, Japan. utetsuya@oita-u.ac.jp.
- J Anesth. 2016 Aug 1; 30 (4): 583-90.
BackgroundDuring epidural anesthesia, the catheter tip occasionally deviates from the epidural space into the intervertebral foramen, resulting in inadequate anesthesia.MethodsDuring postoperative plain radiography, iohexol was injected via the epidural catheter to determine its position and to observe the spread of the material. After exclusion of seven patients with catheters that migrated into the subcutaneous area and 25 patients with no evidence of the contrast medium, 415 patients were evaluated. We retrospectively compared patients to determine whether the incidence of deviation into the intervertebral foramen differed between four types of epidural catheters. We also investigated the load applied to the catheter tip using a Shimadzu Autograph AG-X-500 N-111 universal testing machine.ResultsDeviation of the epidural catheter into the intervertebral foramen was observed in eight and 33 patients in the Hakko and Perifix Soft tip catheter groups, respectively. The incidence of deviation was higher in the Perifix Soft tip catheter group, and lower in the FlexTip Plus and Perifix FX catheter groups. A rapid increase was observed in the force exerted on the tips of the Hakko and Perifix Soft tip catheters, while the force transmitted to the tips of the FlexTip Plus and Perifix FX catheters gradually increased and then reached a plateau at a low level.ConclusionsThe incidence of deviation was significantly lower with spiral-type catheters than with other types of catheters. This might be attributable to the gradual transmission of a lower level of force to the tip in spiral-type catheters.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.