Intensive care medicine · Jan 1993
Comparative StudyPerformance of a hydrophobic heat and moisture exchanger at different ambient temperatures.
- M Croci, A Elena, and M Solca.
- Servizio di Anestesia e Rianimazione, IRCCS Ospedale Maggiore, Milano, Italy.
- Intensive Care Med. 1993 Jan 1; 19 (6): 351-2.
ObjectiveTo evaluate the effect of different room temperatures on hydrophobic heat and moisture exchangers (HME) humidifying capability and efficiency.MethodsStock HMEs were tested in vitro using an already described test model, with separation of inspiratory and expiratory gas. Absolute humidity (AH) was measured by means of dry-wet dual thermocouple, and HME efficiency was computed as the ratio between expired to inspired AH, at room temperature of 20 and 26 degrees C.ResultsInspired gas temperature and AH were significantly higher at 26 than at 20 degrees C; since expired AH remained substantially unchanged, HME efficiency was also higher in warmer environment.ConclusionsHydrophobic HMEs appear to be affected by room temperature, increasing their humidifying ability and their efficiency with its rise.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.