• J Orthop Trauma · May 2014

    Metaphyseal screw augmentation of the LISS-PLT plate with polymethylmethacrylate improves angular stability in osteoporotic proximal third tibial fractures: a biomechanical study in human cadaveric tibiae.

    • Michael Goetzen, Tomas Nicolino, Ladina Hofmann-Fliri, Michael Blauth, and Markus Windolf.
    • *Biomedical Services, AO Research Institute Davos, Davos, Switzerland; and †Department of Trauma Surgery, Medical University of Innsbruck, Innsbruck, Austria.
    • J Orthop Trauma. 2014 May 1; 28 (5): 294-9.

    ObjectivesThe incidence of osteoporotic proximal tibial fractures has increased during the last 2 decades. A promising approach in osteoporotic fracture fixation is polymethylmethacrylate-based cement augmentation of implants to gain better implant purchase in the bone. This study investigates the biomechanical benefits of screw augmentation in less invasive stabilization system-proximal lateral tibial (LISS-PLT) plates in cadaveric extraarticular comminuted proximal tibial fractures (OTA-41-A3.3).MethodsStandardized extraarticular proximal tibial fractures were stabilized with the LISS-PLT plate in 6 paired osteoporotic cadaveric tibiae. Bone mineral density was measured with high-resolution, quantitative computed tomography scans to identify bone quality. In the augmented group, the 5 proximal screws of the LISS-PLT plate were augmented with 1 mL of bone cement each, whereas the contralateral tibia was instrumented conventionally as the control. Cyclic axial loading was applied to each specimen with a starting load of 150 N, using a ramp of 0.05 N per cycle to 10-mm axial displacement. Varus displacement was identified from anterior-posterior radiographs.ResultsBone mineral density showed no significant difference between the 2 groups (P = 0.47). The nonaugmented group reached 9417 load cycles (SD 753) until failure, compared with 14,792 load cycles (SD 2088) in the augmented group (P = 0.002). In the early-onset failure (deformation at 8250 load cycles), varus displacement was significantly smaller in the augmented group (0.46 degrees, SD 0.6) than in the nonaugmented group (3.23 degrees, SD 1.7) (P = 0.01).ConclusionsThis biomechanical study showed that cement augmentation of the LISS-PLT plate screws in osteoporotic proximal extraarticular tibial fractures significantly lowers the propensity toward screw migration and secondary varus displacement.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.