- J Frøkjaer and B N Møller.
- Department of Orthopedic Surgery, Haderslev Hospital, Denmark.
- Acta Orthop Scand. 1992 Aug 1; 63 (4): 434-6.
AbstractIn a prospective study, 25 displaced ankle fractures were treated with one or two biodegradable rods (Biofix), sizes 3.2 or 4.5 mm. In 2 cases the medial malleolus fractured during drilling. At the 1-year follow-up of 22 cases, the following complications had occurred: sinus formation in 1 case and osteolysis corresponding to the fracture line and/or drill channels in 9 cases, 2 of whom had symptoms. 2 patients had been reoperated on, one for pseudarthrosis and another for secondary fracture displacement. Because of the high complication rate and the unknown importance of osteolysis formation, we cannot recommend biodegradable fixation with rods of 3.2 or 4.5 mm length as a supplement to the treatment of displaced ankle fractures, particularly of medial fractures.