Ann Cardiol Angeiol (Paris) · Feb 2016
Comparative Study[Validation during exercise of a new device for cardiac output measurement using pulse wave transit time (comparison EsCCO(®) vs. Physioflow(®))].
- A Stalter, N Lanot, G Bridon, V Julian, B Péreira, and R Richard.
- Service de médecine du sport et des explorations fonctionnelles, hôpital G.-Montpied, CHRU de Clermont-Ferrand, 58, rue Montalembert, 63003 Clermont-Ferrand cedex, France; INRA, UMR 1019, 63000 Clermont-Ferrand, France.
- Ann Cardiol Angeiol (Paris). 2016 Feb 1; 65 (1): 1-6.
ObjectivesEsCCO is a novel non-invasive continuous cardiac output monitoring system based on pulse wave transit time already validated at rest. The aim of our study was to compare cardiac output measurements obtained simultaneously by EsCCO(®) (Q˙cOP) and impedance cardiography (Physioflow(®) ; Q˙cIMP), in healthy subjects.Patients And MethodsEight healthy subjects (age: 31±9 years, weight: 76±10kg, height: 179±5cm) realized two exercise tests: an incremental ergocycle test performed until exertion (Pmax=269±48W) and a constant load exercise (P=163±27W). Comparison between measurements (Q˙cOP versus Q˙cIMP) obtained during the first test allowed to evaluate the accuracy of the device. Reliability was determined on three repeated measures during the second test, realized at ventilatory threshold.ResultsCorrelation coefficient between both methods is 0.88 (P<0.01). Mean difference is 0.04±1.49L/min (95 % limits of agreement: +2.94 to -3.00L/min) and only 3/74 measures are not included between the limits of agreement. At high intensity and for cardiac output over than 15 L/min, Q˙cOP signal is lost in almost half the time. Concerning reliability, reproducibility coefficient is 0.87 (P<0.05), only 1.8 % of this variability is due to the method.ConclusionEsCCO(®) measurements are accurate, reliable and allow a good estimation of cardiac output on healthy subjects. The signal lost observed for high cardiac output levels (>15L/min) can limit its utilization during very high intensity exercise.Copyright © 2015 Elsevier Masson SAS. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.