• Anaesthesia · Mar 1991

    Review Comparative Study

    Potential errors in pulse oximetry. II. Effects of changes in saturation and signal quality.

    • R K Webb, A C Ralston, and W B Runciman.
    • Department of Anaesthesia and Intensive Care, Royal Adelaide Hospital, South Australia.
    • Anaesthesia. 1991 Mar 1; 46 (3): 207-12.

    AbstractThe published studies of pulse oximeter performance under conditions of normal, high and low saturation, exercise, poor signal quality and cardiac arrhythmia are reviewed. Most pulse oximeters have an absolute mean error of less than 2% at normal saturation and perfusion; two-thirds have a standard deviation (SD) of less than 2%, and the remainder an SD of less than 3%. Some pulse oximeters tend to read 100% with fractional saturations of 97-98%. Pulse oximeters may be suitable hyperoxic alarms for neonates if the alarm limit chosen is directly validated for each device. Pulse oximeters are poorly calibrated at low saturations and are generally less accurate and less precise than at normal saturations; nearly 30% of 244 values reviewed were in error by more than 5% at saturations of less than 80%. Ear, nose and forehead probes respond more rapidly to rapid desaturation than finger probes, but are generally less accurate and less precise. Ear oximetry may be inaccurate during exercise. Low signal quality can result in failure to present a saturation reading, but data given with low signal quality warning messages are generally no less accurate than those without. Cardiac arrhythmias do not decrease accuracy of pulse oximeters so long as saturation readings are steady.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.