• Artificial organs · Jul 2013

    Impact of pump replacement on outcome in advanced heart failure patients with left ventricular assist system.

    • Yoshio Iwashima, Masanobu Yanase, Takeshi Horio, Osamu Seguchi, Yoshihiro Murata, Tomoyuki Fujita, Koichi Toda, Yuhei Kawano, and Takeshi Nakatani.
    • Division of Hypertension and Nephrology, National Cerebral and Cardiovascular Center, Suita City, Osaka, Japan. iwashima@hsp.ncvc.go.jp
    • Artif Organs. 2013 Jul 1; 37 (7): 606-14.

    AbstractPump thrombosis is one of the major adverse events associated with the use of a left ventricular assist system (LVAS) in patients with advanced heart failure. We investigated the clinical implication of pump replacement because of thrombus formation. This study included 87 patients who underwent implantation of a Nipro (Toyobo) pulsatile extracorporeal LVAS intended as a bridge to transplantation and were alive more than 3 months after implantation. The pump of this device is translucent, and daily evaluation for signs of thrombus formation was performed. Pump replacement was performed for significant thrombus formation that became visible. Data collection including demographics as well as hematologic values were performed 1 day before (baseline) and 3 months after implantation, and all patients were followed for 2 years or until death. At 3 months after LVAS implantation, 41 patients (47.1%) had undergone pump replacement because of pump thrombus. Baseline body surface area <1.63 m(2) was a significant predictor of pump replacement (hazard ratio [HR] 2.15, P = 0.04). At 3 months after implantation, there was a significantly higher incidence of stroke (P < 0.05) as well as a significantly greater decrease in body weight (F = 4.92, P = 0.03) in patients who underwent pump replacement as compared to those without. The 2-year mortality after implantation was 26.4%. Multivariate Cox regression analysis showed that pump replacement within 3 months after implantation was an independent predictor of mortality (HR 2.50, P = 0.03). In conclusion, pump replacement for thrombus formation may have a strong association with worse outcome. Our results reconfirm the clinical importance of device thrombus in the management of LVAS.© 2013, Copyright the Authors. Artificial Organs © 2013, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…