• Curr Med Res Opin · May 2013

    Comparative Study Clinical Trial

    Accuracy and preference assessment of prefilled insulin pen versus vial and syringe with diabetes patients, caregivers, and healthcare professionals.

    • Andreas Pfützner, Timothy Bailey, Carlos Campos, Douglas Kahn, Ellen Ambers, Marcus Niemeyer, German Guerrero, David Klonoff, and Irina Nayberg.
    • ICRD Institute, San Jose, CA, USA. AndreasP@ikfe.de
    • Curr Med Res Opin. 2013 May 1; 29 (5): 475-81.

    ObjectivesThe primary objective of this study was to investigate the dosing accuracy of the new prefilled FlexTouch insulin pen (FT) in comparison to conventional vial and syringe (V&S) when used by patients (Pts), caregivers (CG) and healthcare professionals (HCPs).MethodsA total of 120 subjects participated in the trial (40 diabetes patients aged 61 ± 11 [mean ± SD] yrs, 20 caregivers [parents and other relatives], 20 physicians, and 40 nurses/certified diabetes educators). The participants were introduced to the devices in randomized order and were asked to perform injections of 5, 25, 43 and 79 IU doses into laboratory tubes. Dosing accuracy was analyzed by weighing the tubes on a pharmaceutical balance and calculating the mean absolute deviation (MAD) from the intended doses. After completing a device assessment questionnaire, Patient Perception Questionnaire (PPQ), with questions regarding device design and performance, the procedure was repeated for the other device, and the patients were finally asked to complete a device preference questionnaire (DPQ).ResultsDosing accuracy was significantly better for FT when used by any of the cohorts at all doses. (MAD ± SD for FT/V&S; 5 IU: 0.4 ± 0.4/0.6 ± 0.6 IU; 25 IU: 0.3 ± 0.4/0.7 ± 0.9 IU; 43 IU: 0.4 ± 0.4/0.9 ± 1.2 IU; 79 IU: 0.5 ± 0.5/1.7 ± 1.6 IU, p < 0.005 for all doses). Dosing accuracy with FT for all three subgroups was comparable (patients: 0.35-0.59 IU; HCP&CG: 0.29-0.54 IU; n.s.). Dosing accuracy with V&S for all three subgroups was not comparable: HCP and CG performed much better with V&S than patients and delivered the doses with significantly higher accuracy (range of mean MAD; patients: 0.81-2.54 IU; HCP&CG: 0.51-1.30 IU, p < 0.005 at all doses). FT was ranked superior to V&S for all aspects of the PPQ. In the DPQ, 93% of the patients voted for FT (neutral: 5%, V&S: 2%), (CG: 100%/0%/0%; HCPs: 85%/2%/13%; p < 0.001 in all cases).ConclusionFT, compared to V&S, was more accurate at all tested doses and was used with similar accuracy by patients, HCPs, and CGs. Using questionnaires only, and without dexterity assessment, study participants rated FT higher than V&S in every component of the PPQ and the vast majority of them preferred FT. These findings may point to a better alternative for dosing accuracy and improved adherence when using the new prefilled insulin pen compared to V&S for insulin delivery in patients with diabetes.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.