J Bone Joint Surg Am · Aug 1997
Randomized Controlled Trial Comparative Study Clinical TrialA comparison of fixation screws for the scaphoid during application of cyclical bending loads.
- E B Toby, T E Butler, T J McCormack, and G Jayaraman.
- The University of Kansas Medical Center, Kansas City 66160, USA.
- J Bone Joint Surg Am. 1997 Aug 1; 79 (8): 1190-7.
UnlabelledMatched pairs of scaphoids from cadavera were stressed with ramped intensity cyclical bending loads after osteotomy and fixation of one scaphoid with a Herbert screw and fixation of the other with an AO 3.5-millimeter cannulated screw, a Herbert-Whipple screw, an Acutrak cannulated screw, or a Universal Compression screw. The AO screw, Acutrak screw, and Herbert-Whipple screw demonstrated superior resistance to cyclical bending loads compared with the Herbert screw. The Universal Compression screw did not provide better fixation than the Herbert screw because of fractures that occurred at the time of insertion. The AO screw and the Herbert screw were then tested in a separate setup in which a segment of volar cortex had been removed in addition to the simple osteotomy. The loss of volar cortex greatly diminished the quality of the fixation provided by both of the screws during application of ramped intensity cyclical bending loads.Clinical RelevanceA fixation device in the scaphoid must be able to withstand the stresses that are placed on the scaphoid as a result of its position spanning the proximal and distal carpal rows. Also, because of the prolonged time required for healing of fractures or non-unions of the scaphoid, the device must be able to withstand many such cycles of stress. The present study demonstrates that commonly used screws for fixation of the scaphoid vary significantly (p < 0.005) in their ability to resist cyclical bending loads.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.