• Eur J Anaesthesiol · Jun 2015

    Comparative Study Clinical Trial Observational Study

    Intraoperative temperature monitoring with zero heat flux technology (3M SpotOn sensor) in comparison with sublingual and nasopharyngeal temperature: An observational study.

    • Timo Iden, Ernst-Peter Horn, Berthold Bein, Ruwen Böhm, Janne Beese, and Jan Höcker.
    • From the Department of Anaesthesiology and Intensive Care Medicine, University Medical Center Schleswig-Holstein, Campus Kiel, Kiel (TI, BB, RB, JB, JH) and Department of Anaesthesiology, Regio Klinikum Pinneberg, Pinneberg, Germany (EPH).
    • Eur J Anaesthesiol. 2015 Jun 1;32(6):387-91.

    BackgroundPerioperative hypothermia is common in patients undergoing general anaesthesia and is associated with important adverse events. The 'gold standard' for monitoring body core temperature - the pulmonary artery catheter - is invasive and unsuitable for most patients. For routine clinical practice, other sites and methods of temperature monitoring are commonly used.ObjectiveThe aim of this study was to evaluate a new temperature sensor (3M SpotOn) using the 'zero heat flux' method attached to the forehead, and compare it to sublingual and nasopharyngeal sensors in terms of correlation, accuracy and precision.DesignAn observational study.SettingUniversity Medical Center Schleswig Holstein, Campus Kiel, Germany from October 2013 to January 2014.PatientsOne hundred and twenty patients scheduled for elective gynaecological or trauma surgery undergoing general anaesthesia were enrolled into this study. Data of 83 patients were finally analysed. Patients with unexpected blood loss, haemodynamic instability determined by the need for continuous norepinephrine infusion and/or need for postoperative ventilation were excluded from this study.InterventionTemperature monitoring was established after induction of anaesthesia with sublingual and nasopharyngeal probes, and the SpotOn sensor.Main Outcome MeasuresBody temperature was measured 15, 45 and 75 min after induction of anaesthesia from sublingual and nasopharyngeal probes and the 3M SpotOn sensor at precisely the same moment.ResultsAnalysis of 83 data sets revealed that 3M SpotOn temperatures were almost identical with nasopharyngeal temperatures (mean difference 0.07 °C; P = 0.1424) and slightly lower than sublingual temperatures by 0.35 °C (P < 0.0001). Coefficients of determination (r) for both methods were between 0.87 (SpotOn vs. nasopharyngeal measurement) and 0.77 (SpotOn vs. sublingual measurement). Bland-Altman analysis revealed a bias (SD) between 0.07 °C (0.21) (SpotOn vs. nasopharyngeal) and -0.35 °C (0.29) (SpotOn vs. sublingual measurement).ConclusionWith respect to correlation, accuracy and precision, the 3M SpotOn sensor provides a good measurement of body temperature in comparison to the nasopharyngeal probe and an acceptable measurement in comparison with sublingual thermometry. It is adequate for clinical use.Trial RegistrationClinicaltrials.gov identifier: NCT02031159.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…