Comparative Study
Flexible laryngoscopy: a comparison of fiber optic and distal chip technologies. Part 1: vocal fold masses.
- Robert Eller, Mark Ginsburg, Deborah Lurie, Yolanda Heman-Ackah, Karen Lyons, and Robert Sataloff.
- USAF Aerodigestive and Voice Center, Wilford Hall Medical Center, San Antonio, TX, USA. robert.eller@lackland.af.mil
- J Voice. 2008 Nov 1; 22 (6): 746-50.
AbstractThis study was designed to evaluate the usefulness of fiber optic (FO) and distal chip (DC) flexible imaging platforms in the diagnosis of true vocal fold pathology when compared to the gold standard rigid transoral laryngeal telescopic examination. The recorded strobovideolaryngoscopic examinations of 34 consecutive patients were evaluated retrospectively by five raters. All stroboscopy segments were evaluated by two laryngologists, an otolaryngologist, a laryngology fellow, and an otolaryngology resident. Seventeen patients were examined with a high-quality, large-diameter, FO flexible laryngoscope (FO group) and 17 random patients were examined with a DC flexible laryngoscope (DC group). Each patient was also examined using rigid laryngeal videostroboscopy at the same sitting. Examinations of three patients from each group were presented twice to monitor internal consistency. Diagnoses of intrinsic vocal fold pathology made with the flexible laryngoscopes were compared for accuracy to the diagnoses provided using the rigid laryngeal telescope. The ability to make clinical diagnoses via stroboscopy was statistically equivalent with FO technology and DC technology. Rigid examination provided more information than the flexible examination in 27% of the FO examinations and in 32% of the DC examinations. DC technology did not add diagnostic information to the examination when compared to a high-quality, large-diameter, FO endoscope. Rigid endoscopy provides superior images of the true vocal folds and is necessary for precise diagnosis in patients with true vocal fold pathology. Thus, the most cost-effective means of evaluation of voice disorders remains FO flexible endoscopy for dynamic voice assessment and the neurolaryngologic examination followed by rigid stroboscopy for evaluation of the vocal fold edge and mucosal wave. Strobovideolaryngoscopy using high-quality FO or DC flexible equipment should be reserved for patients who cannot tolerate transoral rigid examination, such as children and those with a very strong gag reflex.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.