Acta Anaesthesiol Scand · Apr 1990
Comparative StudyComparison of anaesthesia ventilators using a lung model.
- A Johnson and M Bengtsson.
- Department of Anesthesiology, University Hospital, Linköping University, Sweden.
- Acta Anaesthesiol Scand. 1990 Apr 1; 34 (3): 222-6.
AbstractThe ventilatory performance of six anaesthesia ventilators (Alphavent Drägerwerk; MCM 801 Ventilator Dameca; Ohmeda 7000 Ventilator and Ohmeda 7800 Ventilator Ohmeda; Servo Ventilator 900 D and Ventilator 710 Siemens-Elema) was tested in a lung model and compared with the ventilatory performance of the old Engström 300 (Gambro Engström). The lung model consisted of one tracheal tube, two bronchial tubes which could be partially occluded, and two lungs (glass jars filled with water to a suitable compliance). Dialled (set) minute volume was compared to recorded minute volume. Gas distribution between obstructed and non-obstructed lung was calculated. All ventilators had a linear relationship between dialled and measured volume. Only the Servo Ventilator 900 D produced a gas distribution which was as good as, or better than that produced by the Engström 300.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.