    BackgroundWBC counts and differentials are currently performed on complex analyzers in either central or satellite laboratories. Rapid and easy to use point of care (POC) CBC testing can benefit certain outpatient clinical settings, where the greatest clinical need has been demonstrated.MethodsWe evaluated a new POC CBC analyzer Chempaq XBC (Chempaq A/S, Denmark) for hemoglobin, leukocyte counts and 3-part differentials and compared the results with established laboratory based Beckman Coulter LH750 analyzers. The performance of these POC analyzers was tested at multiple clinic locations as a part of regular care on both venous and finger stick blood samples.ResultsMethod validation parameters including precision, accuracy and linearity studies are within the acceptable limits between POC and lab based analyzers. Method comparison studies at various locations showed good correlation at various clinical settings including ER, primary care, Ob/Gyn, ICU, Pedi clinic, Hematology-Oncology clinics, and in-patient ward's in venous blood samples. In addition, at the hematology-oncology clinic, the comparisons of venous as well as finger stick blood sample analyses showed good correlation.ConclusionThe Chempaq XBC analyzer provides accurate hematologic results that can facilitate rapid quantitative assessment of CBC parameters and thus is clinically relevant, especially in outreach clinic settings and in critically ill patients.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.