An automated system for testing the accuracy of patient-controlled analgesia devices.
- D J Hawkins, D R Roberts, A H Ilsley, J L Plummer, and H Owen.
- Department of Anaesthesia and Intensive Care, Flinders Medical Centre, Bedford Park, South Australia.
- Anaesthesia. 1992 Aug 1; 47 (8): 693-6.
AbstractA system was developed to test the accuracy of patient-controlled analgesia devices in situations simulating clinical use. Bolus requests are made automatically at predetermined intervals, and the infusate delivered is measured and recorded without the need for operator presence. To ensure clinical relevance, the bolus request times used in this study corresponded to a pattern typical of those requested by patients on the ward. Graseby, Abbott Provider 5500 and IVAC patient-controlled analgesia devices were tested and found to deliver reasonably accurately over a 24 h period. However, when an infusion was started in an unprimed system or after a period of no bolus requests in a bolus-only mode the Graseby and IVAC machines under-delivered. This system provides a means of testing patient-controlled analgesia devices operating in any delivery mode.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.