AbstractNoninfected unwanted pacemaker leads are usually abandoned since the reported complication rate related to them is low. We followed 60 patients with noninfected retained leads, and complication was observed in 12 (20%) of them. Lead migration occurred in 5 patients, skin erosion in 3 patients, venous thrombosis in 2 patients, and muscle stimulation in 2 patients. Management of the complications was a surgical procedure in seven patients, including two cases of open heart surgery, while chronic medical treatment was necessary in the other five patients. The results of this study suggest that complications due to noninfected abandoned leads may not be as rare as it was previously thought and may present a significant morbidity and cost burden. With the lead extraction technique available, the issue of the removal of all unwanted pacemaker leads should be addressed.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.