Vascular guide wire navigation with a magnetic guidance system: experimental results in a phantom.
- Mirko Schiemann, Reinmar Killmann, Martin Kleen, Nasreddin Abolmaali, Jennifer Finney, and Thomas J Vogl.
- Institute of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Johann Wolfgang Goethe-University, Theodor-Stern-Kai 7, D-60590 Frankfurt am Main, Germany. m.schiemann@vff.uni-frankfurt.de
- Radiology. 2004 Aug 1; 232 (2): 475-81.
PurposeTo investigate the efficacy of a second-generation prototype magnetic guidance system in complex vessel phantoms versus conventional navigation in simulated interventional radiology procedures and to analyze procedure and fluoroscopy times.Materials And MethodsThe magnetic guidance system consists of two focused-field permanent magnets on each side of the body that create a 0.1-T navigation field and is integrated with a modified C-arm single-planar digital angiography system. Forty-nine navigations in a glass phantom and 80 navigations in a three-dimensional liver phantom were performed with a magnetically tipped floppy 0.014-inch guide wire and a conventional 0.014-inch microcatheter system. Rates of success and fluoroscopy and procedure times were quantified for both techniques. For the liver phantom experiment, the Mann-Whitney U test was used. For the glass phantom experiment, the Wilcoxon matched pair test was used with the Hodges-Lehmann estimator.ResultsIn the glass phantom experiments, 42 of 49 turns were successfully performed with both methods. Procedure time to reach a target did not differ significantly between methods, while fluoroscopy time was significantly different when compared with that of the magnetic guidance system (P <.01). Navigation in the liver phantom was successful in 80 of 80 turns with the magnetic guidance system and in 76 of 80 turns with conventional navigation. With the support of the magnetic guidance system, procedure time and fluoroscopy time were significantly different from those with conventional navigation (P <.001).ConclusionThe magnetic guidance system allows the precise navigation of a magnetic guide wire in complex vessel phantoms with significantly shorter fluoroscopy and procedure times.Copyright RSNA, 2004
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.