Comparative Study
Comparison of neonatal skin sensor temperatures with axillary temperature: does skin sensor placement really matter?
- Dorothea Schafer, Sheri Boogaart, Lynette Johnson, Catherine Keezel, Liga Ruperts, and Karen J Vander Laan.
- Neonatal Services, Helen DeVos Children's Hospital, Spectrum Health Medical Center, Grand Rapids, Michigan (Mss Schafer, Boogaart, Johnson, Keezel, and Ruperts); and Nursing Practice & Development, Spectrum Health Hospitals, Grand Rapids, Michigan (Dr Vander Laan).
- Adv Neonatal Care. 2014 Feb 1; 14 (1): 52-60.
PurposeAppropriate thermoregulation affects both morbidity and mortality in the neonatal setting. Nurses rely on information from temperature sensors and radiant warmers or incubators to appropriately maintain a neonate's body temperature. Skin temperature sensors must be repositioned to prevent skin irritation and breakdown. This study addresses whether there is a significant difference between skin sensor temperature readings from 3 locations on the neonate and whether there is a significant difference between skin sensor temperatures compared with digital axillary temperatures.SubjectsThe study participants included 36 hemodynamically stable neonates, with birth weight of 750 g or more and postnatal age of 15 days or more, in a neonatal intensive care unit. Gestational age ranged from 29.6 to 36.1 weeks at the time of data collection.DesignA method-comparison design was used to evaluate the level of agreement between skin sensor temperatures and digital axillary thermometer measurements.MethodsWhen the neonate's skin sensor was scheduled for routine site change, 3 new skin sensors were placed-1 each on the right upper abdomen, left flank, and right axilla. The neonate was placed in a supine position and redressed or rewrapped if previously dressed or wrapped. Subjects served as their own controls, with temperatures measured at all 3 skin sensor sites and followed by a digital thermometer measurement in the left axilla. The order of skin sensor temperature measurements was randomly assigned by a computer-generated number sequence.Main Outcome MeasuresAn analysis of variance for repeated measures was used to test for statistical differences between the skin sensor temperatures. The difference in axillary and skin sensor temperatures was calculated by subtracting the reference standard temperature (digital axillary) from the test temperatures (skin temperatures at 3 different locations), using the Bland-Altman method. The level of significance was set at P < .05.Principal ResultsNo statistically significant differences were found between skin temperature readings obtained from the 3 sites (F2,70 = 2.993, P = .57). Differences between skin temperature readings and digital axillary temperature were also not significant when Bland-Altman graphs were plotted.ConclusionsFor hemodynamically stable neonates in a supine position, there were no significant differences between skin sensor temperatures on abdomen, flank, or axilla or between skin sensor temperatures and a digital axillary temperature. This may increase nurses' confidence that various sites will produce accurate temperature readings.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.