Nitrogen washout/washin, helium dilution and computed tomography in the assessment of end expiratory lung volume.
- Davide Chiumello, Massimo Cressoni, Monica Chierichetti, Federica Tallarini, Marco Botticelli, Virna Berto, Cristina Mietto, and Luciano Gattinoni.
- Dipartimento di Anestesia, Rianimazione (Intensiva e Subintensiva) e Terapia del Dolore, Fondazione IRCCS - Ospedale Maggiore Policlinico Mangiagalli Regina Elena, via Francesco Sforza 35, 20122, Milano, Italy. chiumello@libero.it
- Crit Care. 2008 Jan 1; 12 (6): R150R150.
IntroductionEnd expiratory lung volume (EELV) measurement in the clinical setting is routinely performed using the helium dilution technique. A ventilator that implements a simplified version of the nitrogen washout/washin technique is now available. We compared the EELV measured by spiral computed tomography (CT) taken as gold standard with the lung volume measured with the modified nitrogen washout/washin and with the helium dilution technique.MethodsPatients admitted to the general intensive care unit of Ospedale Maggiore Policlinico Mangiagalli Regina Elena requiring ventilatory support and, for clinical reasons, thoracic CT scanning were enrolled in this study. We performed two EELV measurements with the modified nitrogen washout/washin technique (increasing and decreasing inspired oxygen fraction (FiO2) by 10%), one EELV measurement with the helium dilution technique and a CT scan. All measurements were taken at 5 cmH2O airway pressure. Each CT scan slice was manually delineated and gas volume was computed with custom-made software.ResultsThirty patients were enrolled (age = 66 +/- 10 years, body mass index = 26 +/- 18 Kg/m2, male/female ratio = 21/9, partial arterial pressure of carbon dioxide (PaO2)/FiO2 = 190 +/- 71). The EELV measured with the modified nitrogen washout/washin technique showed a very good correlation (r2 = 0.89) with the data computed from the CT with a bias of 94 +/- 143 ml (15 +/- 18%, p = 0.001), within the limits of accuracy declared by the manufacturer (20%). The bias was shown to be highly reproducible, either decreasing or increasing the FiO2 being 117+/-170 and 70+/-160 ml (p = 0.27), respectively. The EELV measured with the helium dilution method showed a good correlation with the CT scan data (r2 = 0.91) with a negative bias of 136 +/- 133 ml, and appeared to be more correct at low lung volumes.ConclusionsThe EELV measurement with the helium dilution technique (at low volumes) and modified nitrogen washout/washin technique (at all lung volumes) correlates well with CT scanning and may be easily used in clinical practice.Trial RegistrationCurrent Controlled Trials NCT00405002.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.