AbstractAnaesthesia machine failure requires rapid solutions to maintain ventilation and anaesthesia. During procedures with poor access to the patient's airway, it may not be possible to use a self-inflating mechanical ventilation device (SIMVD) for emergency ventilation, and alternative solutions are needed. We evaluated five methods for rescue ventilation using a patient simulator. In Method 1, we used the inspiratory and expiratory tubes and the alternative common gas outlet (ACGO) on the anaesthesia machine to produce a Mapleson E system. In Method 2, we used the tubes, ACGO and an open-ended reservoir bag to produce a Mapleson F system, controlling the bag to achieve ventilation. In Method 3, we attached a SIMVD to the inspiratory tube, and controlled occlusion of the expiratory tube. In Method 4, we used the tubes and ACGO in a Mapleson F configuration, replacing the open-ended bag with a SIMVD to facilitate manual ventilation. In Method 5, we attached a SIMVD to the expiratory tube and left the inspiratory tube attached to its mounting. We were able to achieve ventilation, maintain inhalational anaesthesia, and prevent expired gas rebreathing in Methods 1 and 2. In Method 3 ventilation was achieved with minimal rebreathing of expiratory gas, but with no inhalational agent. Methods 4 and 5 led to rebreathing. Our findings indicate that Methods 1 or 2 are the preferred rapid solutions to maintain ventilation and inhalational anaesthesia in the event of anaesthesia machine failure where there is poor airway access.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.