- S Dugani, A Kumar, and A R Wilkes.
- Department of Anaesthetics and Intensive Care Medicine, School of Medicine, Cardiff University, Cardiff, UK.
- Anaesthesia. 2010 May 1; 65 (5): 468-72.
AbstractWe measured the level of contamination in 207 breathing system filters of five different models. The median (IQR [range]) levels of contamination measured in relative light units on the machine side of the filters were: HMEF 750/S 27 (16-56 [4-13 615]); Ultipor BB25 26 (13-40 [9-66]); Humid-vent filter pedi 19 (15-34 [11-48]); Hygroboy 11 (7-19 [3-113]); and Hygrobaby 9 (6-14 [4-21]). A total of 41/138 (30%) of the HMEF 750/S (the most commonly used filter) had measured values > 50, indicating excessive contamination on the machine side of the filter. The incidence of coughing and duration of the case were significantly associated with the incidence of excessive contamination on the machine side (p = 0.034 and p = 0.024, respectively). Excessive contamination on the machine side of the filter could be from the patient or from the re-used breathing system and could result in cross-infection. The level of contamination may need to be checked routinely during each list.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.