- Eileen Youshock Evanina and Jill L Hanisak.
- Wyoming Valley Health Care System/University of Scranton School of Nurse Anesthesia, Wilkes-Barre, PA, USA.
- AANA J. 2005 Apr 1; 73 (2): 111-3.
AbstractThe electromyographic endotracheal tube (EMG-ETT) is a relatively new tool used to measure integrity of the vocal cord structures during surgery. We describe a case in which an EMG-ETT was inserted for the operative period but not replaced with an ETT during the immediate postoperative period. Intensive care unit nurses had difficulty suctioning the EMG-ETT. The patient was not provided the pulmonary toilet necessary until the EMG-ETT was removed and replaced with a regular ETT. The purpose of this article is to make anesthesia providers aware that when mechanical ventilation is required during the postoperative period, the EMG-ETT should be removed and replaced with a regular ETT to facilitate pulmonary toilet.