-
- Tina A Leone, Allison Lange, Wade Rich, and Neil N Finer.
- Department of Pediatrics, Division of Neonatology, University of California, 200 W Arbor Dr, MC 8774, San Diego, California 92103, USA. tleone@ucsd.edu
- Pediatrics. 2006 Jul 1; 118 (1): e202-4.
AbstractColorimetric carbon dioxide detectors are useful indicators of proper endotracheal tube placement. We have found that they also are helpful during bag and mask ventilation as an indicator of a patent airway. In this report, we describe our experience with these devices for use during preintubation airway stabilization as observed during videotaped performances from a prospective, randomized trial of intubation premedication.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.