    Effect of calibration and environmental condition on the performance of direct-reading organic vapor monitors.

    • Christopher Coffey, Ryan LeBouf, Larry Lee, James Slaven, and Stephen Martin.
    • National Institute for Occupational Safety and Health, 1095 Willowdale Road, Morgantown, WV 26505, USA. ccc3@cdc.gov
    • J Occup Environ Hyg. 2012 Jan 1; 9 (11): 670-80.

    AbstractThe performance of three MIRAN SapphIRe Portable Infrared Ambient Air Analyzers and three Century Portable Toxic Vapor Analyzers equipped with photoionization (PID) and flame ionization (FID) detectors was compared with charcoal tube sampling. Relationships were investigated using two different calibration methods at four cyclohexane concentrations, three temperatures, and four relative humidities. For the first method, the TVA monitors were calibrated with a single concentration of methane for the FID, and isobutylene for the PID. The SapphIRe monitors were zeroed and the monitor's manufacturer-supplied library was used. For the second method, a five-point cyclohexane calibration curve was created for each monitor. Comparison of the monitor results of each calibration method (pooled data) indicated a significant difference between methods (t-test, p < 0.001), The SapphIRe group had results closer to the charcoal tubes with the second calibration method, while the PID and FID monitor groups performed better using the first calibration method. The PID monitor group's performance was affected only at the 90% relative humidity (RH) condition. Using the first method, the monitor readings were compared with the charcoal tube average using mixed linear model analyses of variance (ANOVAs) and regression. The ANOVA results showed there was a statistically significant difference among readings from all monitor types (p <0.0001). The regression results demonstrated that the SapphIRe (r² = 0.97) and FID (r² = 0.92) monitor groups correlated well with the charcoal tubes. The PID monitor group had a similar correlation when 90% RH was excluded (r² = 0.94) but had a weaker correlation when it was included (r² = 0.58). The operator should take care when using these monitors at high concentrations and the PID monitors at high humidities, consider the variability between units of the same monitor, and conduct performance verification of the monitor being used.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…