Anaesth Intensive Care · May 2015
Review Meta AnalysisSystematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7™ or Pronto-7™) and HemoCue® absorption spectrometers (B-Hemoglobin or 201+) with laboratory haemoglobin estimation.
- R Hiscock, D Kumar, and S W Simmons.
- Department of Anaesthesia, Mercy Hospital for Women, Heidelberg, Victoria.
- Anaesth Intensive Care. 2015 May 1;43(3):341-50.
AbstractWe assessed agreement in haemoglobin measurement between Masimo pulse co-oximeters (Rad-7™ and Pronto-7™) and HemoCue® photometers (201+ or B-Hemoglobin) with laboratory-based determination and identified 39 relevant studies (2915 patients in Masimo group and 3084 patients in HemoCue group). In the Masimo group, the overall mean difference was -0.03 g/dl (95% prediction interval -0.30 to 0.23) and 95% limits of agreement -3.0 to 2.9 g/dl compared to 0.08 g/dl (95% prediction interval -0.04 to 0.20) and 95% limits of agreement -1.3 to 1.4 g/dl in the HemoCue group. Only B-Hemoglobin exhibited bias (0.53, 95% prediction interval 0.27 to 0.78). The overall standard deviation of difference was larger (1.42 g/dl versus 0.64 g/dl) for Masimo pulse co-oximeters compared to HemoCue photometers. Masimo devices and HemoCue 201+ both provide an unbiased, pooled estimate of laboratory haemoglobin. However, Masimo devices have lower precision and wider 95% limits of agreement than HemoCue devices. Clinicians should carefully consider these limits of agreement before basing transfusion or other clinical decisions on these point-of-care measurements alone.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.