Internally Randomized Control Trial of Radiation Exposure Using Ultra-low Radiation Imaging Versus Traditional C-arm Fluoroscopy for Patients Undergoing Single-level Minimally Invasive Transforaminal Lumbar Interbody Fusion.
- Timothy Y Wang, S Harrison Farber, Scott S Perkins, Adam G Back, Sarah A Byrd, Debbie Chi, David Vincent, and Isaac O Karikari.
- Department of Neurological Surgery, Duke University Medical Center, Durham, NC.
- Spine. 2017 Feb 15; 42 (4): 217223217-223.
Study DesignRandomized controlled trial.ObjectiveTo compare radiation exposure between ultra-low radiation imaging (ULRI) with image enhancement and standard-dose fluoroscopy for patients undergoing minimally invasive transforaminal lumbar interbody fusion (MIS TLIF).Summary Of Background DataAlthough the benefits of MIS are lauded by many, there is a significant amount of radiation exposure to surgeon and operating room personnel. Our goal with this work was to see if by using ultra-low dose radiation settings coupled with image enhancement, this exposure could be minimized.MethodsAn institutional review board approved, prospective, internally randomized controlled trial was performed comparing ultra-low dose settings coupled with image enhancement software to conventional fluoroscopic imaging. In this study, each patient served as their own control, randomly assigning one side of MIS-TLIF for cannulation and K-wire placement using each imaging modality. Further, the case was also randomly divided into screw placement and cage placement/final images to allow further comparisons amongst patients. Radiation production from the C-arm fluoroscope and radiation exposure to all operating room personnel were recorded.ResultsTwenty-four patients were randomly assigned to undergo a single level MIS-TLIF. In no case was low radiation imaging abandoned, and no patient had a neurologic decline or required hardware repositioning. Everyone in the operating room-the physician, scrub nurse, circulator, and anesthesiologist-all benefited with 61.6% to 83.5% reduction in radiation exposure during cannulation and K-wire placement to screw insertion aided by ULRI. In every case but the anesthesiologist dose, this was statistically significant (P < 0.05). This benefit required no additional time (P = 0.78 for K-wire placement).ConclusionULRI, when aided by image enhancement software, affords the ability for all parties in the operating room to substantially decrease their radiation exposure compared with standard-dose C-arm fluoroscopy without adding additional time or an increased complication rate.Level Of Evidence2.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.