Response of 10 pulse oximeters to an in vitro test system.
- K J Reynolds, J T Moyle, M K Sykes, and C E Hahn.
- Department of Anaesthetics, John Radcliffe Hospital, Oxford.
- Br J Anaesth. 1992 Apr 1; 68 (4): 365-9.
AbstractPulse oximeters are often used in situations in which severe hypoxaemia may occur. We have developed an in vitro system to test the accuracy of pulse oximeter calibration. The probe of 10 different oximeters was attached to a model finger in an in vitro blood circuit, and pulse oximeter readings (SpO2) were compared with multi-wavelength in vitro oximeter readings (SO2) over a range of SO2 values from 50 to 100%. The oximeters tested varied widely in their accuracy and linearity. We conclude that the system can test the accuracy, reproducibility and linearity of response of pulse oximeter readings at low oxyhaemoglobin saturations.