Prediction of flow capability in intravenous infusion systems: implications for fluid resuscitation.
- B K Philip and J H Philip.
- Department of Anesthesia, Brigham and Women's Hospital, Harvard Medical School, Boston, MA 02115.
- J Clin Monit. 1990 Apr 1; 6 (2): 113-7.
AbstractThe pressure-flow (P-F) relationship for intravenous infusion systems is nonlinear and may be expressed by the quadratic model P = RLF + RTF2. The flow parameters RL and RT may represent the resistance of laminar and turbulent flow, respectively. In this study pressure and flow were measured, and RL and RT were calculated for several infusion tubings, catheter, and system components. We then developed a technique to identify the relative effect of various devices on achieving the higher flows needed for fluid resuscitation. A typical infusion system was chosen, and the experimentally determined flow parameters RL and RT of its components were used in the quadratic P-F relationship at P = 300 mm Hg. Devices in the infusion system were ranked, using a subtractive algorithm, according to their relative impediment to flow as measured by contribution to the pressure drop. The order of devices removed or replaced, from largest to least pressure drop, was as follows: fluid warmer, 16-gauge catheter, check valve, 14-gauge catheter, standard-bore Y tubing, 12-gauge catheter, and standard-bore stopcock, leaving 10-gauge catheter + wide tubing. Devices with large RT, such as fluid warmers and check valves that produce large pressure drops, should generally be avoided during fluid resuscitation when high flows are needed. A similar ordering of device removal or substitution (largest to least pressure drop) was determined using the traditional but incorrect linear P-F model, P = RF, and the order of devices chosen for elimination was different.(ABSTRACT TRUNCATED AT 250 WORDS)
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.