• Can J Anaesth · Nov 1991

    Evaluation of oxygen saturation monitoring by pulse oximetry in neonates in the delivery system.

    • I Dimich, P P Singh, A Adell, M Hendler, N Sonnenklar, and M Jhaveri.
    • Department of Anesthesiology, Mount Sinai Medical School, New York, New York.
    • Can J Anaesth. 1991 Nov 1; 38 (8): 985-8.

    AbstractThe pulse oximeter was evaluated for use in neonates in the delivery room. One hundred neonates, delivered vaginally or by Caesarean section with general or epidural anaesthesia, were studied. After delivery, pulse oximetry probes were placed simultaneously on the ulnar side of the right hand and on the right Achilles tendon to determine whether there was a difference in arterial oxygenation (SpO2). Measurements of SpO2 were taken at 1, 5, 10 min, and 24 hr after delivery. At one and five minutes, SpO2 recorded from the right hand was higher than that recorded from the lower extremities (71.9% +/- 6.5% vs 63.4% +/- 4.3% and 83.3% +/- 4.2% vs 76% +/- 4.1%, mean +/- SD, respectively). At ten minutes these differences diminished, and had almost completely disappeared after 24 hr. These results can be explained by the presence of R-L shunting at the ductus arteriosus level, producing reduced SaO2 in the lower extremities. Oxygen saturation did not differ between neonates delivered vaginally or by Caesarean section, regardless of the presence or type of anaesthesia. We concluded that neonates remain relatively desaturated in the immediate postpartum period and that the SpO2 obtained from the right hand is a better index of neonatal oxygenation than that obtained from the heel.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.