• ASAIO J. · Nov 2007

    Comparative Study Clinical Trial

    The effect of temperature correction of blood gas values on the accuracy of end-tidal carbon dioxide monitoring in children after cardiac surgery.

    • Pertti K Suominen, Stephen Stayer, Wei Wang, and Anthony C Chang.
    • Department of Anesthesia and Intensive Care, Hospital for Children and Adolescents, Helsinki University Central Hospital, Helsinki, Finland.
    • ASAIO J. 2007 Nov 1; 53 (6): 670-4.

    AbstractWe evaluated accuracy of end-tidal carbon dioxide tension (PETco2) monitoring and measured the effect of temperature correction of blood gas values in children after cardiac surgery. Data from 49 consecutive mechanically ventilated children after cardiac surgery in the cardiac intensive care unit were prospectively collected. One patient was excluded from the study. Four arterial-end-tidal CO2 pairs in each patient were obtained. Both the arterial carbon dioxide tension (Paco2) values determined at a temperature of 37 degrees C and values corrected to body temperature (Patcco2) were compared with the PETco2 values. After the surgical correction 28 patients had biventricular, acyanotic (mean age 2.7 +/- 4.8 years) and 20 patients had a cyanotic lesion (mean age 1.0 +/- 1.7 years). The body temperature ranged from 35.2 degrees C to 38.9 degrees C. The Pa-PETco2 discrepancy was affected both by the type of cardiac lesion and by the temperature correction of Paco2 values. Correlation slopes of the Pa-PETco2 and Patc-PETco2 discrepancies were significantly different (p = 0.040) when the body temperature was higher or lower than 37 degrees C. In children, after cardiac surgery, end-tidal CO2 monitoring provided a clinically acceptable estimate of arterial CO2 value, which remained stabile in repeated measurements. End-tidal CO2 monitoring more accurately reflects temperature-corrected blood gas values.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…