• J. Am. Coll. Cardiol. · Sep 2014

    Multicenter Study Clinical Trial

    Transcatheter aortic valve replacement for severe symptomatic aortic stenosis using a repositionable valve system: 30-day primary endpoint results from the REPRISE II study.

    • Ian T Meredith Am, Darren L Walters, Nicolas Dumonteil, Stephen G Worthley, Didier Tchétché, Ganesh Manoharan, Daniel J Blackman, Gilles Rioufol, David Hildick-Smith, Robert J Whitbourn, Thierry Lefèvre, Rüdiger Lange, Ralf Müller, Simon Redwood, Dominic J Allocco, and Keith D Dawkins.
    • MonashHEART, Monash Health, and Monash University, Clayton, Victoria, Australia. Electronic address: ian.meredith@myheart.id.au.
    • J. Am. Coll. Cardiol. 2014 Sep 30; 64 (13): 1339-48.

    BackgroundTranscatheter aortic valve replacement provides results comparable to those of surgery in patients at high surgical risk, but complications can impact long-term outcomes. The Lotus valve, designed to improve upon earlier devices, is fully repositionable and retrievable, with a unique seal to minimize paravalvular regurgitation (PVR).ObjectivesThe prospective, single-arm, multicenter REPRISE II study (REpositionable Percutaneous Replacement of Stenotic Aortic Valve Through Implantation of Lotus Valve System: Evaluation of Safety and Performance) evaluated the transcatheter valve system for treatment of severe symptomatic calcific aortic valve stenosis.MethodsPatients (n = 120; aortic annulus 19 to 27 mm) considered by a multidisciplinary heart team to be at high surgical risk received the valve transfemorally. The primary device performance endpoint, 30-day mean pressure gradient, was assessed by an independent echocardiographic core laboratory and compared with a pre-specified performance goal. The primary safety endpoint was 30-day mortality. Secondary endpoints included safety/effectiveness metrics per Valve Academic Research Consortium criteria.ResultsMean age was 84.4 years, 57% of the patients were female, and 76% were New York Heart Association functional class III/IV. Mean aortic valve area was 0.7 ± 0.2 cm(2). The valve was successfully implanted in all patients, with no cases of valve embolization, ectopic valve deployment, or additional valve implantation. All repositioning (n = 26) and retrieval (n = 6) attempts were successful; 34 patients (28.6%) received a permanent pacemaker. The primary device performance endpoint was met, because the mean gradient improved from 46.4 ± 15.0 mm Hg to 11.5 ± 5.2 mm Hg. At 30 days, the mortality rate was 4.2%, and the rate of disabling stroke was 1.7%; 1 (1.0%) patient had moderate PVR, whereas none had severe PVR.ConclusionsREPRISE II demonstrates the safety and effectiveness of the Lotus valve in patients with severe aortic stenosis who are at high surgical risk. The valve could be positioned successfully with minimal PVR. (REPRISE II: REpositionable Percutaneous Replacement of Stenotic Aortic Valve Through Implantation of Lotus™ Valve System - Evaluation of Safety and Performance; NCT01627691).Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…