- J Herlitz, A Bång, A Axelsson, J R Graves, and J Lindqvist.
- Division of Cardiology, Sahlgrenska University Hospital, Göteborg, Sweden.
- Resuscitation. 1998 Apr 1; 37 (1): 3-7.
AimTo describe the sequences of arrhythmias, number of shocks delivered and the number of failures in a consecutive series of patients with out-of-hospital cardiac arrest attended by our emergency medical service (EMS) and in whom cardio-pulmonary resuscitation (CPR) was initiated and in whom automated external defibrillators (AEDs) were used.PatientsAll patients with out-of-hospital cardiac arrest attended by the EMS and in whom AEDs were used. Time for inclusion in the study: January 1st, 1987 to December 31st, 1992.ResultsIn all there were 1781 out of hospital cardiac arrests during the study period. Among them AEDs were used in 383 cases (22%). The total number of interpreted rhythms delivered in these patients was 2719. Among all rhythm sequences coarse ventricular fibrillation (VF) was found on 375 occasions (14%); fine VF on 107 occasions (4%) and ventricular tachycardia (VT) on 12 occasions (0.4%). In ten cases with coarse VF (nine patients) the AED did not advise a shock (2.7%). In five of those nine patients a human error was interpreted as the explanation and in four there was a possible technical error. In these four patients defibrillation was delayed by 33-43 s, respectively. Among the 2225 rhythm sequences not judged as VF/VT the AED advised a shock on one occasion (0.04%).ConclusionAmong patients with coarse VF AED gave inaccurate instructions in 2.7%. However, the majority of the failures were judged to be caused by human errors.