Intensive care medicine · Mar 2002
Clinical TrialEvaluation of a new device for noninvasive measurement of nonshunted pulmonary capillary blood flow in patients with acute lung injury.
- Marcelo Gama de Abreu, Stefan Geiger, Tilo Winkler, Max Ragaller, Thomas Pfeiffer, Dirk Leutheuser, and Detlev Michael Albrecht.
- Clinic of Anesthesiology and Intensive Care Medicine, University Clinic Carl Gustav Carus, Technical University Dresden, Fetscherstrasse 74, 01307, Dresden, Germany. mgabreu@aol.com
- Intensive Care Med. 2002 Mar 1; 28 (3): 318-23.
ObjectivesTo evaluate the performance of a new device for noninvasive measurement of nonshunted pulmonary capillary blood flow (PCBF) by partial CO2 rebreathing.Design And SettingProspective clinical trial in an intensive care unit of a university hospital.Patients And ParticipantsTwenty mechanically ventilated patients with acute lung injury.InterventionsVariations in PEEP of +/-3 cmH2O.Measurements And ResultsInitially PCBF was measured invasively as cardiac output minus venous admixture (Q(VA)/Q(t)) flow, and by partial CO2 rebreathing at baseline PEEP (PEEP(b)). The PEEP was then reduced by 3 cmH2O (to PEEP(b-3)) and measurements were repeated after 30 min. PEEP was then increased by 6 cmH2O (to PEEP(b+3)), and measurements were repeated after 10, 20, and 30 min. The overall correlation coefficient between noninvasive and invasive PCBF measurements at PEEP(b) was high ( r=0.97), with close agreement between methods being observed (0.1+/-0.6 l/min, bias and precision, respectively). Accordingly, both the correlation coefficient and agreement between methods for changes in PCBF from PEEP(b-3) to PEEP(b+3) levels were satisfactory ( r=0.71; 0.2+/-0.5 l/min, bias and precision). The new device was able to detect the correct PCBF trend in 17 of 20 patients investigated and in all patients who showed invasive PCBF changes equal to or greater than 0.3 l/min ( n=12). Noninvasive PCBF changes were stable as early as 10 min after variation in PEEP, as compared to 30 min values.ConclusionsThe new device appears to be clinically useful for the monitoring of PCBF in patients suffering from acute lung injury. Our results suggest that titration of PEEP aimed at improving PCBF can be performed with the new device.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.