• J Magn Reson Imaging · Jan 2001

    Metallic surgical instruments for interventional MRI procedures: evaluation of MR safety.

    • F G Shellock.
    • Department of Radiology, University of Southern California, School of Medicine, Los Angeles, California 90045, USA. www.MRIsafety.com
    • J Magn Reson Imaging. 2001 Jan 1; 13 (1): 152-7.

    AbstractThis investigation evaluated metallic surgical instruments for magnetic resonance (MR) safety in association with a 1.5-Tesla/64-MHz MR system. Seven different instruments (mallet, bone punch, curette, Weil-Blakesley ethmoid forceps, suction cannula, septum speculum, and Kocher-Langenbeck retractor; Aesculap, Inc. (South San Francisco, CA) were tested for magnetic field interactions, heating, and generation of artifacts by using previously described techniques. Heating was evaluated for the septum speculum and Kocher-Langenbeck retractor by using a special gel-filled phantom and a fluoroptic thermometer to record temperatures immediately before and during MRI performed at a whole-body averaged SAR of 1.3 W/kg. Artifacts were assessed with the instruments placed inside of a gel-filled phantom and performing MRI using T1-weighted spin-echo and gradient-echo pulse sequences. Magnetic field interactions were relatively minor (deflection angles, 0 to 7 degrees; torque, 0 to +1), the highest temperature changes were < or = +0.8 degrees C, and the artifacts should not create substantial problems considering the "intended use" for these instruments. The findings of the MR safety tests indicated that the seven different metallic surgical instruments (Aesculap, Inc.) would be safe and acceptable for use in interventional MRI procedures performed with MR systems with static magnetic fields of 1.5 T or less. J. Magn. Reson. Imaging 2001;13:152-157.Copyright 2001 Wiley-Liss, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.