• Clinical biomechanics · May 2009

    Factors influencing interlocking screw failure in unreamed small diameter nails--a biomechanical study using a distal tibia fracture model.

    • Patrick Weninger, Michael Schueller, Michael Jamek, Stefanie Stanzl-Tschegg, Heinz Redl, and Elmar K Tschegg.
    • Trauma Center "Lorenz Boehler", Donaueschingenstrasse 13, 1200 Vienna, Austria. patrick.weninger@gmx.net
    • Clin Biomech (Bristol, Avon). 2009 May 1; 24 (4): 379-84.

    BackgroundUnreamed tibia nails with small diameters are increasingly used for fracture fixation. However, little is known about the fatigue strength of proximal and distal interlocking screws in those nails. To date, no data are available reporting on mechanical differences of solid compared to cannulated tibial nails. The aim of this study was to assess the fatigue strength of proximal and distal interlocking screws of solid and cannulated small diameter tibia nails.MethodsWe created a distal tibia fracture model (AO/OTA 43 A3) using 16 Sawbones. After fracture stabilization with one of four different nail types (Expert Tibial Nail, VersaNail, T2 Tibial Nailing System, Connex), mechanical testing was performed in three loading series (40,000 cycles each) with incremental loads. Timing and type of interlocking screw failure were assessed.FindingsInterlocking screw failure was observed significantly earlier (after a mean interval of 57,042 cycles) in cannulated tibial nails (VersaNail, T2) compared to solid nails (after a mean interval of 88,415 cycles; P < 0.001). Proximal interlocking screw failure was recorded if oblique screws were used proximally (VersaNail, T2, Connex). No distal interlocking screw failure was recorded in the Connex nail. Two- and three-part fractures of proximal or distal interlocking screws were observed in all specimen.InterpretationProximal and distal interlocking screw failure has to be considered in small diameter nails in case of delayed fracture healing. To support our results, further experimental studies and clinical series are necessary.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…