• Scand J Trauma Resus · Jan 2015

    Quality of basic life support when using different commercially available public access defibrillators.

    • Michael P Müller, Cynthia Poenicke, Maxi Kurth, Torsten Richter, Thea Koch, Carolin Eisold, Adrian Pfältzer, and Axel R Heller.
    • Department of Anaesthesiology and Intensive Care Medicine, University Hospital, Technische Universität Dresden, Dresden, Germany. mpmueller.web@gmail.com.
    • Scand J Trauma Resus. 2015 Jan 1;23:48.

    BackgroundBasic life support (BLS) guidelines focus on chest compressions with a minimal no-flow fraction (NFF), early defibrillation, and a short perishock pause. By using an automated external defibrillator (AED) lay persons are guided through the process of attaching electrodes and initiating defibrillation. It is unclear, however, to what extent the voice instructions given by the AED might influence the quality of initial resuscitation.MethodsUsing a patient simulator, 8 different commercially available AEDs were evaluated within two different BLS scenarios (ventricular fibrillation vs. asystole). A BLS certified instructor acted according to the current European Resuscitation Council 2010 Guidelines and followed all of the AED voice prompts. In a second set of scenarios, the rescuer anticipated the appropriate actions and started already before the AED stopped speaking. A BLS scenario without AED served as the control. All scenarios were run three times.ResultsThe time until the first chest compression was 25 ± 2 seconds without the AED and ranged from 50 ± 3 to 148 ± 13 seconds with the AED depending on the model used. The NFF was .26 ± .01 without the AED and between .37 ± .01 and .72 ± .01 when an AED was used. The perishock pause ranged from 12 ± 0 to 46 ± 0 seconds. The optimized sequence of actions reduced the NFF, which ranged now from .32 ± .01 to .41 ± .01, and the perishock pause ranging from 1 ± 1 to 19 ± 1 seconds.ConclusionsVoice prompts given by commercially available AED merely meet the requirements of current evidence in basic life support. Furthermore, there is a significant difference between devices with regard to time until the first chest compression, perishock pause, no-flow fraction and other objective measures of the quality of BLS. However, the BLS quality may be improved with optimized handling of the AED. Thus, rescuers should be trained on the respective AED devices, and manufacturers should expend more effort in improving user guidance to shorten the NFF and perishock pause.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…