Randomized Controlled Trial Comparative Study
Comparison study of upper arm and forearm non-invasive blood pressures in adult Emergency Department patients.
- Karen Schimanski, Andrew Jull, Nancy Mitchell, and Jessica McLay.
- Emergency Department, Auckland City Hospital, Private Bag 92024, Auckland, New Zealand. Electronic address: karen.schimanski@adhb.govt.nz.
- Int J Nurs Stud. 2014 Dec 1; 51 (12): 1575-84.
BackgroundForearm blood pressures have been suggested as an alternative site to measure blood pressures when the upper arm is unavailable. However there is little evidence utilising clinical populations to support this substitution.ObjectivesTo determine agreement between blood pressures measured in the left upper arm and forearm using a singular oscillometric non-invasive device in adult Emergency Department patients. The secondary objective was to explore the relationship of blood pressure differences with age, sex, ethnicity, smoking history and obesity.DesignSingle centre comparison study.SettingAdult Emergency Department, Tertiary Trauma Centre.ParticipantsForty-four participants who met inclusion/exclusion criteria selected sequentially from the Emergency Department arrival board.MethodsA random assignment of order of measurement for left upper arm and forearm blood pressures was utilised. Participants were eligible if they were aged 18 years or older, had been assigned an Australasian Triage Scale code of 2, 3, 4, or 5, were able to consent, and able to have blood pressures measured on their left arm whilst lying at a 45° angle. The Bland-Altman method of statistical analysis was used, with the level of agreement for clinical acceptability for the systolic, diastolic and mean arterial pressure defined as ±10 mmHg.ResultsThe forearm measure overestimated systolic (mean difference 2.2 mmHg, 95% limits of agreement ±19 mmHg), diastolic (mean difference 3.4 mmHg, 95% limits of agreement ±14.4 mmHg), and mean arterial pressures (mean difference 4.1 mmHg, 95% limits of agreement ±13.7 mmHg). The systolic measure was not significantly different from zero. Evidence of better agreement was found with upper arm/forearm systolic measures below 140 mmHg compared to systolic measures above 140 mmHg using the Levene's test (p=0.002, F-statistic=11.09). Blood pressure disparity was not associated with participant characteristics.ConclusionsForearm measures cannot routinely replace upper arm measures for blood pressure measurement. If the clinical picture requires use of forearm blood pressure, the potential variance from an upper arm measure is ±19 mmHg for systolic pressure, although the variability may be close to ±10 mmHg if the systolic blood pressure is below 140 mmHg.Copyright © 2014 Elsevier Ltd. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.