Critical care medicine · Feb 1983
Comparative StudyEffect of the dynamic response of transducer-tubing system on accuracy of direct blood pressure measurement in patients.
- A Boutros and S Albert.
- Crit. Care Med. 1983 Feb 1; 11 (2): 124-7.
AbstractThe dynamic responses of 3 transducer-tubing systems (TTS) were documented using square and sine wave generators. The TTS tested were P23 ID with 5 feet tubing (P23-5'), P23 ID with 6 inches tubing (P23-6") and P50A transducer with no tubing (P50). Natural frequencies were 45, 33, and 7 Hz for P50, P23-6", and P23-5', respectively. All were underdamped (damping coefficients 0.1-0.3). Simultaneous measurements of blood pressures by the 3 TTS revealed that systolic pressures recorded by P23-5' were significantly higher (p less than 0.001) than by P23-6" (17%) and P50 (16%). This difference was not influenced by the size of the arterial cannula or by the heart rate. However, the differences were greater with radial, compared to brachial arterial cannulation (p less than 0.02). There were no differences in systolic pressures measured with P23-6" and P50. There were no differences in diastolic or electronic mean pressures recorded by the 3 TTS.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.