• Clin. Orthop. Relat. Res. · Sep 2015

    Cathodic Electrical Stimulation Combined With Vancomycin Enhances Treatment of Methicillin-resistant Staphylococcus aureus Implant-associated Infections.

    • Scott Nodzo, Menachem Tobias, Lisa Hansen, Nicole R Luke-Marshall, Ross Cole, Linda Wild, Anthony A Campagnari, and Mark T Ehrensberger.
    • Department of Orthopedics, State University of New York at Buffalo, Buffalo, NY, USA.
    • Clin. Orthop. Relat. Res. 2015 Sep 1; 473 (9): 2856-64.

    BackgroundEffective treatments for implant-associated infections are often lacking. Cathodic voltage-controlled electrical stimulation has shown potential as a treatment of implant-associated infections of methicillin-resistant Staphylococcus aureus (MRSA).Questions/PurposesThe primary purpose of this study was to (1) determine if cathodic voltage-controlled electrical stimulation combined with vancomycin therapy is more effective at reducing the MRSA bacterial burden on the implant, bone, and synovial fluid in comparison to either treatment alone or no treatment controls. We also sought to (2) evaluate the histologic effects of the various treatments on the surrounding bone; and to (3) determine if the cathodic voltage-controlled electrical stimulation treatment had an effect on the mechanical properties of the titanium implant as a result of possible hydrogen embrittlement.MethodsThirty-two adult male Long-Evans rats (Harlan Laboratories, Indianapolis, IN, USA) with surgically placed shoulder titanium implants were infected with a clinical strain of MRSA (NRS70). One week after infection, eight animals received a treatment of cathodic voltage-controlled electrical stimulation at -1.8 V versus Ag/AgCl for 1 hour (STIM), eight received vancomycin twice daily for 1 week (VANCO), eight received the cathodic voltage-controlled electrical stimulation and vancomycin therapy combined (STIM + VANCO), and eight served as controls with no treatment (CONT). Two weeks after initial infection, the implant, bone, and synovial fluid were collected for colony-forming unit (CFU) enumeration, qualitative histological analysis by a pathologist blinded to the treatments each animal received, and implant three-point bend testing.ResultsThe implant-associated CFU enumerated from the STIM + VANCO (mean, 3.7 × 10(3); SD, 6.3 × 10(3)) group were less than those from the CONT (mean, 1.3 × 10(6); SD, 2.8 × 10(6); 95% confidence interval [CI] of difference, -4.3 × 10(5) to -9.9 × 10(3); p < 0.001), STIM (mean, 1.4 × 10(6); SD, 2.0 × 10(6); 95% CI of difference, -2.1 × 10(6) to -1.8 × 10(3); p = 0.002), and VANCO (mean, 5.8 x 10(4); SD, 5.7 × 10(4); 95% CI of difference, -6.4 × 10(4) to -1.7 × 10(4); p < 0.001) group. The bone-associated CFU enumerated from the STIM + VANCO group (6.3 × 10(1); SD, 1.1 × 10(2)) were less than those from the CONT (mean, 2.8 × 10(5); SD, 4.8 × 10(5); 95% CI of difference, -9.4 × 10(4) to -5.0 × 10(3); p < 0.001) and STIM (mean, 2.6 × 10(4); SD, 2.5 × 10(4); 95% CI of difference, -4.1 × 10(4) to -1.6 × 10(3); p < 0.001) groups. The VANCO group (4.3 × 10(5); SD, 6.3 × 10(2)) also had lower bone-associated CFU as compared with the CONT (mean 95% CI of difference, -9.3 × 10(4) to -4.5 × 10(3); p < 0.001) and STIM (95% CI of difference, -4.0 × 10(4) to -1.5 × 10(3); p < 0.001) groups. In comparison to the synovial fluid CFU enumerated from the CONT group (mean, 3.3 × 10(4); SD, 6.0 × 10(4)), lower synovial CFU were reported for both the STIM + VANCO group (mean, 4.6 × 10(1); SD, 1.2 × 10(2); 95% CI of difference, -4.9 × 10(3) to -3.0 × 10(2); p < 0.001) and the VANCO group (mean, 6.8 × 10(1); SD, 9.2 × 10(1); 95% CI of difference, -4.9 × 10(3) to -2.8 × 10(2); p = 0.007). The histological analysis showed no discernable deleterious effects on the surrounding tissue as a result of the treatments. No brittle fracture occurred during mechanical testing and with the numbers available, no differences in implant flexural yield strength were detected between the groups.ConclusionsIn this rodent model, cathodic voltage-controlled electrical stimulation combined with vancomycin is an effective treatment for titanium implant-associated infections showing greater than 99.8% reduction in bacterial burden on the implant, surrounding bone, and synovial fluid as compared with the controls and the stimulation alone groups.Clinical RelevanceCathodic voltage-controlled electrical stimulation combined with vancomycin may enable successful treatment of titanium orthopaedic implant-associated infections with implant retention. Future studies will focus on optimization of the stimulation parameters for complete eradication of infection and the ability to promote beneficial host tissue responses.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…