-
Randomized Controlled Trial Comparative Study
A Randomized Crossover Study Comparing a Novel Needle Guidance Technology for Simulated Internal Jugular Vein Cannulation.
- David B Auyong, Stanley C Yuan, Alyse N Rymer, Cynthia L Green, and Neil A Hanson.
- From the Department of Anesthesiology, Virginia Mason Medical Center, Seattle, Washington (D.B.A., S.C.Y., A.N.R., N.A.H.); and Department of Biostatistics, Duke Clinical Research Institute, Durham, North Carolina (C.L.G.).
- Anesthesiology. 2015 Sep 1;123(3):535-41.
BackgroundDespite ultrasound guidance for central line placement, complications persist, as exact needle location is often difficult to confirm with standard two-dimension ultrasound. A novel real-time needle guidance technology has recently become available (eZono, Germany) that tracks the needle during insertion. This randomized, blinded, crossover study examined whether this needle guidance technology improved cannulation of a simulated internal jugular (IJ) vein in an ultrasound phantom.MethodsOne hundred physicians were randomized to place a standard needle in an ultrasound neck phantom with or without the needle guidance system. Video cameras were placed externally and within the lumens of the vessels to record needle location in real time. The primary outcome measured was the rate of posterior wall puncture. Secondary outcomes included number of carotid artery punctures, number of needle passes, final needle position, time to cannulation, and comfort level with this new technology.ResultsThe incidence of posterior vessel wall puncture without and with needle guidance was 49 and 13%, respectively (P < 0.001, odds ratio [OR] = 7.33 [3.44 to 15.61]). The rate of carotid artery puncture was higher without needle navigation technology than with needle navigation 21 versus 2%, respectively (P = 0.001, OR = 12.97 [2.89 to 58.18]). Final needle tip position being located within the lumen of the IJ was 97% accurate with the navigation technology and 76% accurate with standard ultrasound (P < 0.001, OR = 10.42 [2.76 to 40.0]). Average time for successful vessel cannulation was 1.37 times longer without guidance technology.ConclusionThis real-time needle guidance technology (eZono) shows significant improvement in needle accuracy and cannulation time during simulated IJ vein puncture.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.