• J. Am. Vet. Med. Assoc. · Dec 2009

    Comparative Study

    Comparison of a human portable blood glucose meter, veterinary portable blood glucose meter, and automated chemistry analyzer for measurement of blood glucose concentrations in dogs.

    • Beth M Johnson, Michael M Fry, Bente Flatland, and Claudia A Kirk.
    • Department of Small Animal Clinical Sciences, College of Veterinary Medicine, University of Tennessee, Knoxville, TN 37996, USA. bethjohnson@utk.edu
    • J. Am. Vet. Med. Assoc. 2009 Dec 1; 235 (11): 1309-13.

    ObjectiveTo compare blood glucose concentrations measured with 2 portable blood glucose meters (PBGMs) validated for use in dogs (PBGM-D) and humans (PBGM-H) and an automated chemistry analyzer.DesignValidation study.Sample Population92 samples of fresh whole blood and plasma from 83 dogs with various diseases.ProceduresEach PBGM was used to measure whole blood glucose concentration, and the automated analyzer was used to measure plasma glucose concentration. Passing-Bablok linear regression and Bland-Altman plots were used to determine correlations and bias between the PBGMs and the automated analyzer. Calculated acceptability limits based on combined inherent instrument imprecision were used with Bland-Altman plots to determine agreement. Clinical relevance was assessed via error grid analysis.ResultsAlthough correlation between results of both PBGMs and the standard analyzer was > 0.90, disagreement was greater than could be explained by instrument imprecision alone. Mean difference between PBGM-H and chemistry-analyzer values was -15.8 mg/dL. Mean difference between PBGM-D and chemistry-analyzer values was 2.4 mg/dL. Linear regression analysis revealed proportional bias of PBGM-H (greater disagreement at higher glucose concentrations); no proportional bias was detected for PBGM-D. No constant bias was detected for either PBGM. Error grid analysis revealed all measurements from both PBGMs were within zones without an anticipated effect on clinical outcome.Conclusions And Clinical RelevanceNeither PBGM had exact agreement with the automated analyzer; however, the disagreement detected did not have serious clinical consequences. Our findings stressed the importance of using the same device for monitoring trends in dogs and using instrument-specific reference ranges.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.