- Christian Veltmann, Juergen Kuschyk, Rainer Schimpf, Florian Streitner, Nina Schoene, Martin Borggrefe, and Christian Wolpert.
- 1st Department of Medicine-Cardiology, University Medical Centre Mannheim, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany. christian.veltmann@umm.de
- Clin Res Cardiol. 2010 Jan 1; 99 (1): 37-44.
BackgroundIn Brugada syndrome implantable cardioverter defibrillator (ICD) therapy is associated with a high rate of inappropriate therapies, mainly due to supraventricular tachyarrhythmias (SVT) (2.7-14.1%/year). Aim of the present study was to evaluate a single ventricular fibrillation (VF) detection zone with a high cut-off rate with respect to prevention of inappropriate ICD shock due to SVT and safety of this programming.MethodsSixty-one consecutive patients (mean age 42.6 +/- 12.9 years; 41 males) diagnosed with Brugada syndrome and implanted with an ICD were included. ICDs were prospectively programmed with a single VF detection zone and a cut-off rate of 222 beats/minute (bpm). A maximum of six shocks with the maximal shock energy were programmed. The minimal follow-up was 1 year.ResultsDuring a follow-up of 47.6 +/- 23.1 months seven patients (2.91%/year) received appropriate ICD shocks. No patient suffered from syncope or died. Five patients (2.07%/year) received inappropriate ICD shocks: four patients due to T-wave oversensing and only one patient (0.4%/year) due to SVT (atrial fibrillation with a ventricular rate of >222 bpm).ConclusionsProgramming of a single, high-rate VF zone in patients with Brugada syndrome and an implanted defibrillator is safe. Such programming may be associated with reduced inappropriate defibrillator discharges. A single detection zone with a high VF cut-off rate can be recommended in patients with Brugada syndrome.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.