J Clin Monit Comput · Aug 2016
Comparative StudyComparison of an advanced minimally invasive cardiac output monitoring with a continuous invasive cardiac output monitoring during lung transplantation.
- Roland Tomasi, Stephan Prueckner, Stephan Czerner, Renè Schramm, Gerhard Preissler, Bernhard Zwißler, and Vera von Dossow-Hanfstingl.
- Department of Anaesthesiology, University of Munich, Marchionistrasse 15, 81377, Munich, Germany.
- J Clin Monit Comput. 2016 Aug 1; 30 (4): 475-80.
AbstractThe aim of this study was to compare a continuous non-calibrated left heart cardiac index (CI) measurement by arterial waveform analysis (FloTrac(®)/Vigileo(®)) with a continuous calibrated right heart CI measurement by pulmonary artery thermodilution (CCOmbo-PAC(®)/Vigilance II(®)) for hemodynamic monitoring during lung transplantation. CI was measured simultaneously by both techniques in 13 consecutive lung transplants (n = 4 single-lung transplants, n = 9 sequential double-lung transplants) at distinct time points perioperatively. Linear regression analysis and Bland-Altman analysis with percentage error calculation were used for statistical comparison of CI measurements by both techniques. In this study the FloTrac(®) system underestimated the CI in comparison with the continuous pulmonary arterial thermodilution (p < 0.000). For all measurement pairs we calculated a bias of -0.55 l/min/m(2) with limits of agreement between -2.31 and 1.21 l/min/m(2) and a percentage error of 55 %. The overall correlations before clamping a branch oft the pulmonary artery (percentage error 41 %) and during the clamping periods of a branch oft the pulmonary artery (percentage error 66 %) failed to reached the required percentage error of less than 30 %. We found good agreement of both CI measurements techniques only during the measurement point "15 min after starting the second one-lung ventilation period" (percentage error 30 %). No agreement was found during all other measurement points. This pilot study shows for the first time that the CI of the FloTrac(®) system is not comparable with the continuous pulmonary-artery thermodilution during lung transplantation including the time periods without clamping a branch of the pulmonary artery. Arterial waveform and continuous pulmonary artery thermodilution are, therefore, not interchangeable during these complex operations.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.