• Surgical endoscopy · Feb 2008

    Comparative Study

    Comparison of four energy-based vascular sealing and cutting instruments: A porcine model.

    • Benjamin Person, David A Vivas, Dan Ruiz, Michael Talcott, James E Coad, and Steven D Wexner.
    • Department of Colorectal Surgery, Cleveland Clinic, 2950 Cleveland Clinic Blvd, Weston, FL 33331, USA. personb@ccf.org
    • Surg Endosc. 2008 Feb 1; 22 (2): 534-8.

    AimTo compare the safety and efficacy of four energy-based vascular sealing and cutting instruments.MethodsBlood vessels of various types and diameters were harvested from four pigs using four instruments: Harmonic ACE (Ethicon Endo-Surgery, Cincinnati, OH), LigaSure V and LigaSure Atlas (Valleylab, Inc., Boulder, CO; a division of Tyco Healthcare), and EnSeal vessel fusion system (SurgRx, Inc. Redwood City, CA). The diameters of the vessels, speed and adequacy of the cutting and sealing process, and bursting pressures were compared. An additional set of specimens was sealed and left in situ for up to 4 h after which the vessels were harvested and histopathologically analyzed for the degree of thermal injury.ResultsThe bursting pressures were significantly higher with EnSeal compared to all other instruments (p < 0.0001). The sealing process was significantly shorter with Harmonic ACE and significantly longer with LigaSure Atlas (p <0.0001). The mean seal width was larger with the LigaSure Atlas compared to the other instruments, and it was smaller with EnSeal and Harmonic ACE. Less radial adventitial collagen denaturation was present with EnSeal and LigaSure V than with the other two instruments; there were no significant differences in collagen denaturation although proximal thermal injury to the smooth muscle in the media of the vessel wall was less common with LigaSure Atlas than with the other instruments; however, the numbers were too small for statistical analysis.ConclusionsThe bursting pressures with EnSeal were significantly higher than with all the other instruments. Harmonic ACE was the fastest sealing instrument and LigaSure Atlas was slowest. EnSeal created less radial thermal damage to the adventitial collagen of the vessels and LigaSure Atlas created less thermal damage to the media of the vessels. The clinical significance of these findings is unknown.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.